Model No.: VKD-3PV-18L 
Dental autoclave 
Key Specifications/Special Features: 
  
has 12L,18L,23L for chose  
  
 
the condition of using the sterilizer: 
Temperature of environment: 5℃~40℃; 
Relative humidity: ≤80%; 
Bactericidal pressure: 86kPa~106kPa; 
Frequency: 50Hz±1Hz,Votage: one-phase 220V±22V. 
Working condition and normal parameters: 
The highest rated working pressure: 0.21~0.23MPa; 
The highest rated temperature of environment: 134~137℃; 
Power: interchange 220V, 50Hz; 
Vapor pressure: 0.3~0.6MPa; 
Water pressure: 0.15~0.3MPa. 
3.2 Process of the sterilizer 
Process sterilizing time Sterilizing temperature Sterilizing pressure 
121℃/unwrapped 30min 121℃ 1.1bar 
121℃/wrapped 30min 121℃ 1.1bar 
134℃/unwrapped 8min 134℃ 2.1bar 
134℃/wrapped 8min 134℃ 2.1bar 
134℃/cotton 8min 134℃ 2.1bar 
 
                     |